Patient Monitor P9000L

P9000L Multi-Para Patient monitor

Features &Benefits:

5 standard parameters: ECG, RESP, NIBP, SPO2, 1-TEMP.

8.4”high resolution color TFT display..

Real time S-T segment analysis, pace-maker detection and ARR analysis.

Multi-display selectable including standard, large font, trend coexis, OxyCRG dynamic

On-line help and Patient info input management

Multi-lead ECG waveforms display in phase.

Large volume of tabular and graphic trends information storage and easy to recall

Capture dynamic waveforms.

Efficient resistance to interference of defibrillator and HF knife.

Up to 4 hours working capacity of built-in rechargeable battery.

Networking capacity and TCP/IP networking platform.

Option of FHR,FM,TOCO,Printer , IBP and EtCO2

Performance Specifications

Display: 8.4’’color TFT

Rolling and refreshing waveform display

Multi display selectable including:

Large-font display

Trend coexist display

OxyCRG dynamic view display.

Bed-to-bed view display

Trace: 9 waveforms(7 ECG, 1 SPO2 and 1 RESP)

Sweep speed:12.5mm/s,25mm/s,50mm/s

Indicator: Power/battery indicator light

QRS beep and alarm sound

Battery: Rechargeable lead acid cell, 12v/4AH

Max.24 hour for charging, 4 hours for continue working

Trend: Parameter graphic and tabular trends:

5s/piece,8 hours;

1min/piece,168 hours(24hour×7days)

5min/piece,1000 hours.

Storage: NIBP:1000 groups

Alarm:200 groups

Full disclosure waveforms:1 hour

Alarm: User-adjustable High, Medium and Low limits 3-level Audible and visual alarm

Networking: Connected to central monitoring system

TCP/IP netting platform

Standard Parameters

ECG:

Lead mode: 5 –lead(R,L,F,N,C)

Lead selection : I,II,III,avR,avL,avF,V

Waveform: 3 and 7 channel selectable

Gain selection: 0.5mm/mv,1mm/mv,2mm/mv

Sweep speed: 12.5mm/s;25mm/s;50mm/s

Heart rate range:

Adult: 15~300bpm;

Neonate:/pediatric:15~350bpm

Accuracy: +1bpm or +1%,whichecer is greater

Resolution: 1bpm

Filter: surgery mode:1~20Hz

monitor model:0.5~40Hz

Diagnostic mode:0.05~130Hz

Scaling signal: 1mv,+3%

Protection: withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation

Alarm range: 15~350bpm

S-T segment detecton:

Measurement range: 2.0mV~+2.0mV

Alarm Range: -2.0mV~ +2.0mV

Accuracy: -0.8mV ~+0.8Mv

Error: +0.02Mv

Arrhythmia analysis: YES

SPO2

Measurement range: 0~100%

Resolution: 1%

Accuracy: +2%(70-100%);0-69% unspecified

Alarm range 0~100%

Pluse Rate: range:20~300bpm

Resolution:1bpm

Error:+1bpm or +2%,whichever is greater

NIBP

Method: Digital Automatic oscillometric

Operation mode : Manual/Automatic/continuous

Auto measurement time: Adjustable(1~480min)

Measurement Unit: mmHg/Kpa selectable

Measurement types: Systolic, Diastolic, Mean

Measurement tange:

Range of Systolic pressure:

Adult: 40~270mmHg

Pediatric: 40~220mmHg

Neonate: 40~135mmHg

Range of mean pressure:

Adult: 20~235mmHg

Pediatric: 20~165mmHg

Neonate: 20~110mmHg

Range of diastolic pressure:

Adult: 10~215mmHg

Pediatric: 10~150mmHg

Neonate :10~100mmHg

Over-pressure protection:

Double safety protection

Resolution: 1 mmHg

Alarm: Systolic.Diastolic,Mean

RESPIRATION

Method: Thoracic impedance

Measurement range: Adult:7~120rpm;

Neonate/Pediatric:7~150rpm

Apnea alarm: YES,10~40s

Resolution: 1rpm

Accuracy: +2rpm

TEMPERATURE

Compatible probe: YSI or CYF

Measurement range: 5~50?

Resolution : 0.1?

Accuracy: +0.1?

Refreshing time: about 1

Average measuring time: <10s

FHR

Transducer: Multi-crystal, Pulsed Doppler

Measurement range: 50~210 BPM

Working frequency: 1 MHz

Strength: <5mW/cm2

Signal processing:

special DSP system and modern recognization.

Resolution: 1BPM

Accuracy: ±1BPM

Alarm Range: High: 160,170,180,190 BPM,

Low: 100,110,120 BPM

FM

Manual button marking,

automatic FM identifying function

TOCO Measuring

Transducer: External pressure transducer

Measurement range: 0~100 units

Resolution: 1rpm

Accuracy: ±2 rpm

IBP

Channel: 2 channels

Range: -50-300mmHg

Resolution: 1mmHg

Accuracy: ±4mmHg(±4%)

Unit: mmHg, Kpa

Converter Sensitivity: 5mV/V/mmHg

Transducer sites: ART/PA/CVP/LAP/RAP/ICP

EtCO2( Sidertream CO2)

Measurement range: 0~99mmHg

Accuracy: +2mmHg (0~40mmHg)

Sampling Range: 100ml/min

Sampling rate accuracy: 15%

Respiration rate: 0~120rmp

Respiration accuracy: +2rmp (0~70rmp)

+5rmp (>70rmp)

Respiration time: <240msec (10% to 90%)

Delay time: <2s

EtCO2 (mainstreaam CO2)

Method: Infrared Spectrum

Range: 0.0-10% (0~76%)

Resolution: 1mmHg (0.1%)

Accuracy: <5%(±4.0 mmHg)

Or<10%(of Readings)

Recorder:

Build-in, thermal array

Plethysmogram caveform:3 channels

Record mode: manual, on alarm, time-defined

Recording width: 50mm

Printing speed: 50mm/s

Recording type: Frozen waveform record

NIBP recall record

Trend table record

Alarm record

Fixed -time record

Miscellaneous

Satety:

Safety level: Class I, type CF

Dimendion and Weight

Dimendion: 440×430×450

G.Weight: <9.0KS

Operation Environment

Temperature: Working 0~+40?

Transportation and storage –20~+60?

Humidity: working=85%

Transportation and storage=93%

Power: AC 100-240,50/60Hz

Patient range:

Neonate, pediatric, and adult patients

Standard Accessories:

(1) 5 lead ECG cable

(2) 1 spo2 probes

(3) 1 NIBP prbe

(4) 1 temp probe

(5)1 ground connecting liner

(6) Chest electrode (10pcs/set)